Monitoring QA EPD plots
Here you can find the most concerning (based on: https://online.star.bnl.gov/epd/ and https://dashboard1.star.bnl.gov/daq/EPD_UI/?EPD )
[most recent on the top]
25 July 2023
In run 24206005 we had some tiny problem with few channels - I guess one card. The run was short, so the fact that later it is ok I would not worry about it:
https://online.star.bnl.gov/epd/Signal/24206/24206005.adc_Rg1_West.png
24 July 2023
Sometimes we are missing QA monitoring plots for some runs, good ones - not yet understood. But I checked them and they seems to be ok, just missing plots. I did the rainbow "scan" - looks ok! - attached TileByTile_24205007.pdf .
15 June 2023
Nothing new
07 June 2023
https://online.star.bnl.gov/epd/Signal/24186/24186018.adc_Rg3_East.png
30 June 2023
No changes
28 June 2023
No changes
27 June 2023
First 5k run was not on the QA plots. Attached ADC plots run: 24177033
Look ok
26 June 2023
No changes.
25 June 2023
No changes.
24 June 2023
No changes.
23 June 2023
After break in printing QA monitoring plots I made some "rainbow plots" - attached. Looks ok. Today also QA monitoring ADC - no surprises.
14 June 2023
Dead guys: WW PP1TT6 and W PP2TT10 still dead. Nothing interesting except of that.
06 June 2023
TILE - 7 pattern info - GUI - online monitoring plots. - links and comments
W PP1 TT6 - Mismatch! (zero on adc plot) - Rmon abnormal, I=0.020muA - high peak at very low adc - https://online.star.bnl.gov/epd/Signal/24157/24157001.adc_Rg2_West.png
W PP2 TT10 - match patterns(OK!) - Rmon abnormal, I=0.020muA - empty - https://online.star.bnl.gov/epd/Signal/24157/24157016.adc_Rg3_West.png
E PP1 TT31 - Mismatch (zero on adc plot) - I=8.330 muA - empty - https://online.star.bnl.gov/epd/Signal/24157/24157012.adc_Rg5_East.png
W PP2 TT27 - Mismatch (zero on adc plot) - I=3.880 muA - adc distribution present, I would say that there is relatively too much dark current? - https://online.star.bnl.gov/epd/Signal/24157/24157014.adc_Rg5_West.png
04 June 2023
DCS Control:
Two tiles W PP2TT10 PP1TT6 has Rmon "abnormal" -> same issue from the beginning
Yesterday I saw the high current in East PP1TT1, not it is higher than yesterday: 10.220 muA - it is in 'grey" which I do not really understand.
ADC
unstable card?
In one run 24155006 we had 4 channels W PP11 TT2,3,4,5 - just dark current
https://online.star.bnl.gov/epd/Signal/24155/24155006.adc_Rg1_West.png
Just dark current also in:
[EDIT ] ->. This is ok !!! W PP2 TT4 and W PP6 TT1 https://online.star.bnl.gov/epd/Signal/24155/24155006.adc_Rg1_West.png
W PP4 TT6 https://online.star.bnl.gov/epd/Signal/24155/24155015.adc_Rg2_West.png
empty :
E PP1 TT31 https://online.star.bnl.gov/epd/Signal/24155/24155016.adc_Rg5_East.png. (same as previous days)
W PP2 TT10 https://online.star.bnl.gov/epd/Signal/24155/24155016.adc_Rg3_West.png (same as previous days)
03 June 2023
DCS Control:
Two tiles W PP2TT10 PP1TT6 has Rmon "abnormal" -> same as yesterday
One tile has high current ~10muA, East PP1TT1, the adc look ok, but has high dark current (top right plot): https://online.star.bnl.gov/epd/Signal/24154/24154001.adc_Rg1_East.png
ADC
Empty:
- SAME AS YESTERDAY: E PP1TT31 cr4 brd0x16 ch7,https://online.star.bnl.gov/epd/Signal/24154/24154006.adc_Rg5_East.png
- SAME AS YESTERDAY: W PP2TT10 cr3 brd0x14 ch14, https://online.star.bnl.gov/epd/Signal/24154/24154006.adc_Rg3_West.png
- W PP2 TT8 CORRECT
Dark current peak SAME AS YESTERDAY: W PP1 TT6 cr2 brd0x18 ch24 https://online.star.bnl.gov/epd/Signal/24154/24154002.adc_Rg2_West.png
02 June 2023
DCS Control:
Two tiles W PP2TT10 PP1TT6 has Rmon "abnormal" as on attached screen
ADC
Empty:
- SAME AS YESTERDAY: E PP1TT31 cr4 brd0x16 ch7, https://online.star.bnl.gov/epd/Signal/24153/24153017.adc_Rg5_East.png
- SAME AS YESTERDAY: W PP2TT10 cr3 brd0x14 ch14, https://online.star.bnl.gov/epd/Signal/24153/24153013.adc_Rg3_West.png
- SAME AS YESTERDAY: W PP2 TT8 cr2 sth wrong with the crate naming https://online.star.bnl.gov/epd/Signal/24153/24153013.adc_Rg2_West.png
Dark current peak SAME AS YESTERDAY: W PP1 TT6 cr2 brd0x18 ch24 https://online.star.bnl.gov/epd/Signal/24153/24153013.adc_Rg2_West.png
01 June 2023
ADC
Empty:
- W PP1 TT6 cr2 brd0x18 ch24 or empty or some dark current https://online.star.bnl.gov/epd/Signal/24152/24152001.adc_Rg2_West.png
- W PP2 TT8 cr2 sth wrong with the crate naming
- W PP2 TT10 cr3 brd0x14 ch14
- E PP1 TT31 cr4 brd0x16 ch7
Dark Current only:
- (Only in run 24152001, then ok: https://online.star.bnl.gov/epd/Signal/24152/24152001.adc_Rg2_West.png)
- W PP3 TT6 cr4 ndr0x18 ch16
- W PP3 TT7 cr4 ndr0x18 ch17
- W PP3 TT8 cr4 ndr0x18 ch18
- W PP3 TT9 cr4 ndr0x18 ch19
- In all runs:
- W PP6 TT1 cr3 ndr0x14 ch3(https://online.star.bnl.gov/epd/Signal/24152/24152008.adc_Rg1_West.png)
- W PP2 TT4 cr3 ndr0x14 ch8
- W PP2 TT4 cr3 ndr0x14 ch8 https://online.star.bnl.gov/epd/Signal/24152/24152017.adc_Rg2_West.png
- mstefaniak's blog