- surrow's home page
- Posts
- 2024
- May (1)
- 2018
- 2017
- 2016
- 2015
- 2014
- October (3)
- September (1)
- August (1)
- July (1)
- June (3)
- May (4)
- April (6)
- March (4)
- February (5)
- January (2)
- 2013
- December (2)
- November (1)
- October (1)
- September (4)
- August (6)
- July (11)
- June (8)
- May (5)
- April (5)
- March (11)
- February (4)
- 2012
- December (3)
- November (3)
- October (3)
- September (3)
- August (3)
- July (10)
- June (9)
- May (5)
- April (4)
- March (4)
- February (7)
- January (4)
- 2011
- 2010
- December (2)
- November (4)
- August (2)
- July (3)
- June (6)
- May (4)
- April (4)
- March (8)
- February (6)
- January (4)
- 2009
- My blog
- Post new blog entry
- All blogs
FGT Commissioning Tasks / Cu-Au running
Trigger / Configuration:
- Low-multiplicity trigger
- A special configuration a al 'fgt_setup' will be prepared
- Data taking will be handed over to FGT crew (Bernd & Gerrit) taking runs under different parameter settings
- Need data available for fast offline processing
Tasks:
- HV:
- Data set 1:
- 90/10 : Perform scan for fixed HV of all 14 quarter sections (2700V - 3700V) with current latency
- Adjust HV to equalize response based on total cluster charge
- Choose mean HV such that efficiency is 'high' based on previous studies (Compare HV voltage values to set 14)
- Data set 2:
- Take high statistics run with new HV setting
- Data set 3:
- Take high statistics run with adjusted HV setting using track reconstruction
- Data set 1:
- Timing:
- Data set 4:
- Perform timing scan with adjusted HV setting
- Data set 4:
- Efficiency scan:
- Data set 5:
- Perform fine HV scan (50V steps) for disk 3 while keep HV for disks 1, 2, 4, 5 and 6 at adjusted values
- Prepate efficiency vs. HV
- Data set 5:
- Gas:
- Change gas to 70/30 and 80/20 and repeat above steps (2 X three data sets plus efficiency scan)
Groups:
- surrow's blog